Oxidative Induction Time (O.I.T.)

  

Analysis of the thermal stability of polymers is important for materials under thermal load such as polymer insulation for electronics. Determination of the oxidative induction time (O.I.T.) is a standardized technique which can be easily carried out using the DSC 200 F3 Maia®. Presented here is a measurement on PE heated to 210°C under inert (nitrogen) conditions. After a five-minute equilibration time, the atmosphere was switched to oxygen; the sample then began to degrade after 70.97 minutes under oxidizing conditions.

  
  back


NETZSCH offers precision thermal analysis instruments including advanced dilatometers (with a new basic R&D/QC dilatometer), classical DSC & TGA (with a new economical DSC), high temperature DSC to 1650°C for specific heat, very high temperature STA (TGA-DSC/DTA) to 2400°C (featuring the new STA 409-PC Luxx®), thermal / evolved gas analysis with fully-integrated FTIR & MS, plus high resolution TMA and DMA. We also feature leading technology for thermal conductivity and diffusivity measurement as well as refractories testing including HMOR, CIC, and RUL.